Data storage

These tables outline the typical data storage needs for CPAP and bilevel devices.

Data type

CPAP devices

Bilevel devices

Patient details

1 KB to 15 KB depending on the number of detailed sessions

Summary therapy data

1 KB per day

1 KB per day

Detailed data

50 KB per day

75 KB per day

High rate data

1.2 MB per day

1.5MB per day

The values in this table are based on eight hours of periodic data per day.

Example calculation

A clinic monitors 5,000 patients over 12-months with follow-ups every three months. Of the 5,000 patients, 4,000 use CPAP devices and the remainder use bilevel devices.

Table 1. Data required at each follow up

Device

Summary therapy

Detailed data

High-rate data

Total per visit

CPAP

1 KB x 91 days

50 KB x 91 days

1.2 MB x 91 days

114 MB

Bilevel

1 KB x 91 days

75 KB x 91days

1.5 MB x 91 days

144 MB



Table 2. Total storage required (with high-rate data)

Number of patients

Number of visits

Data size

Storage per year

CPAP patients

4,000

4

114 MB

1824 GB

Bilevel patients

1,000

4

144 MB

576 GB

Total per year

2400 GB



Table 3. Total storage required (with no high-rate data)

Number of patients

Number of visits

Data size

Storage per year

CPAP patients

4.000

4

4.6 MB

73.6 GB

Bilevel patients

1,000

4

6.9 MB

27.62 GB

Total per year

101.2 GB