These tables outline the typical data storage needs for CPAP and bilevel devices.
Data type | CPAP devices | Bilevel devices |
|---|---|---|
Patient details | 1 KB to 15 KB depending on the number of detailed sessions | |
Summary therapy data | 1 KB per day | 1 KB per day |
Detailed data | 50 KB per day | 75 KB per day |
High rate data | 1.2 MB per day | 1.5MB per day |
The values in this table are based on eight hours of periodic data per day.
Example calculation
A clinic monitors 5,000 patients over 12-months with follow-ups every three months. Of the 5,000 patients, 4,000 use CPAP devices and the remainder use bilevel devices.
Device | Summary therapy | Detailed data | High-rate data | Total per visit |
|---|---|---|---|---|
CPAP | 1 KB x 91 days | 50 KB x 91 days | 1.2 MB x 91 days | 114 MB |
Bilevel | 1 KB x 91 days | 75 KB x 91days | 1.5 MB x 91 days | 144 MB |
Number of patients | Number of visits | Data size | Storage per year | |
|---|---|---|---|---|
CPAP patients | 4,000 | 4 | 114 MB | 1824 GB |
Bilevel patients | 1,000 | 4 | 144 MB | 576 GB |
Total per year | 2400 GB |
Number of patients | Number of visits | Data size | Storage per year | |
|---|---|---|---|---|
CPAP patients | 4.000 | 4 | 4.6 MB | 73.6 GB |
Bilevel patients | 1,000 | 4 | 6.9 MB | 27.62 GB |
Total per year | 101.2 GB |